6 Operation and maintenance

894 Professional CVS

■■■■■■■■

75

3 Place the measuring head cover and the stopper (3-5) onto the mea-

suring head.

4 Start the determination and wait for the result.

If the curve meets the following conditions, then defects in the hard-
ware or the electrode cables can be ruled out:

The curve must be linear.

At –2.0 V, the current should be –1.67 mA ± 0.33 mA.

At +2.0 V, the current should be +1.67 mA ± 0.33 mA.

Figure 47

Dummy cell test – Ideal curve progression

5 If the resulting curve is not consistent with this representation and

the problems with the determinations persist, contact the local
Metrohm Service.