Figure 16 – Metrohm 894 Professional CVS User Manual

Page 34

Advertising
background image

3.2 Equipping the RDE measuring head

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

894 Professional CVS

3 Allow the bridge electrolyte to react in the electrolyte vessel until the

diaphragm is soaked with bridge electrolyte.

4 Place the reference electrode in the filled electrolyte vessel and

tighten it.

The electrolyte solution that is displaced in the electrolyte vessel is
forced out of the deaeration openings.

Figure 16

Assembling the reference electrode with the electrolyte
vessel

5 Rinse the assembled reference electrode.

6 Insert the installed reference electrode into the opening (5-20) of the

measuring head insert.

7

CAUTION

The electrode cables for the reference and the auxiliary electrode
are identical. Observe the markings on the plugs, because the two
cables must not be mixed up.

Advertising