8 Appendix

8.1

Tubing lengths in the measuring head arm

The following list provides information about the lengths and diameters of
the tubings used in the measuring head arm between the connector plate
and the tubing connection port. You will need this information for setting
the parameters of the dosing units in viva (e.g. for the function Prepare).

Connection
between con-
nector plate
(1-4)

… and tubing
connector port
of the measur-
ing head arm
(see Figure 6,
page 15)

Article number

Tubing length
in mm

Diameter in
mm

Connector 1

Connector 1

6.1805.550

250

2.00

Connector 2

Connector 2

6.1805.550

250

2.00

Connector 3

Connector 3

6.1805.550

250

2.00

Connector SMP

Connector SMP

6.1831.020

220

0.75

Connector
WASTE

Connector
WASTE

6.1805.050 and

6.1805.540

210 and

80

2.00

Connector OUT

Connector OUT

6.1805.550

250

2.00

8.2

"Status" LED – Various instrument statuses

The "Status" LED on the front of the instrument indicates three different
statuses:

Continuously on

The instrument is ready for operation.

Blinking regularly

The instrument is operating.

Blinking pattern
"LED on a long
time - off a short
time - on a long
time - off a short
time…"

A standby potential is being applied to the

electrodes. Do not remove the measuring head
or the electrode cables in this instrument status.