3 galvanostat, 4 temperature measurement, 5 measuring input – Metrohm 894 Professional CVS User Manual

Page 97: 6 accuracy, Galvanostat, Temperature measurement, Measuring input, Accuracy

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

9 Technical specifications

894 Professional CVS

■■■■■■■■

87

9.3

Galvanostat

Potential measur-
ing ranges

–5.00 - +5.00 V

9.4

Temperature measurement

Measuring range
(Pt1000)

0 - +100 °C

Accuracy (Pt1000)

±0.5 °C

9.5

Measuring input

Measuring interval

Sampling rate

100 kHz

Input impedance
(RE)

> 15 G

Ω // < 8 pF

Input bias current
(RE) at 25 °C

< 2 pA

Bandwidth of the
electrometer

25 MHz

9.6

Accuracy

Current applied

±(0.2% of the current +0.2% of the current measuring range)

Current measured

±(0.2% of the current +0.2% of the current measuring range)

Potential applied

±(0.2% of the potential ± 1 mV)

Potential mea-
sured

±(0.2% of the potential ± 1 mV)