5 signatures, Electronic signature – Metrohm tiBase 1.0 Manual User Manual

Page 96

Advertising
background image

3.2 Administration

■■■■■■■■■■■■■■■■■■■■■■

88

■■■■■■■■

tiBase 1.0

[Set]

Sets the parameters to the default settings.

Comment on modification of determinations

If this check box is activated then each time that a determination is modified
a modification reason and a modification comment must be entered; these
are saved in the determination and shown in the database in the subwindow
Information on the tab Determination.

Selection

on | off

Default value

off

3.2.2.5

Signatures

Tab: Configuration

Tools Security settings Signatures

On the tab Signatures the parameters for the electronic signatures can be
configured.

Recommended settings

[Set]

Sets the parameters to the default settings.

Inactivity delay

on | off (Default value: off)

If this check box is activated then the dialog window for signing will be
automatically closed when the entered time limit has expired.

Range

1 ... 60 min

Default value

10 min

Remove password after signature

on | off (Default value: off)

If this check box is activated then the password must be entered again after
each signature.

Sign each determination separately

on | off (Default value: off)

If this check box is activated then each determination contained in the
determination overview must be signed individually.

Advertising