10 Appendix

899 Coulometer

■■■■■■■■

99

3 Confirming the initialization

Note

If the request is not confirmed within 8 s, then the procedure will
be canceled.

Press [BACK] twice.

Initialization is started. The process takes approximately 80 s. The
instrument will be automatically restarted after successful initializa-
tion.

10.7

Remote interface

10.7.1

Pin assignment of the remote interface

1

5

6

9

5

1

9

6

Figure 27

Pin assignment of remote socket and remote plug

The above figure of the pin assignment applies for all Metrohm instru-
ments with 9-pin D-Sub remote connector.

Table 8

Inputs and outputs of the remote interface

Pin No.

Assignment

Function

1

Output 0

Ready/EOD

2

Output 1

-

3

Output 2

Titration

4

Output 3

Cond OK

5

Output 4

Error

6

0 volt (GND)

7

+5 volts

8

Input 0

Start

9

Input 1

Stop