7 Parameters

899 Coulometer

■■■■■■■■

75

Selection

off

Default value

off

Measured value display

If this parameter is activated, then the currently measured value is dis-
played during the conditioning.

Selection

on | off

Default value

off

7.1.2

Start conditions

Menu

Parameters Start conditions

The parameters that are carried out before the start of titration are
defined under Start conditions.

Pause

Waiting time, for example for dissolving the sample. During this time no
iodine is generated.

Input range

0 ... 999999 s

Default value

0 s

Request sample ID

Selection of the sample identification that is queried in the method run.

Selection

ID1 | ID2 | ID1&ID2 | off

Default value

off

Request sample size

If this parameter is activated, then the value for the sample size will be
requested.

Selection

on | off

Default value

on

Request sample unit

If this parameter is activated, then the unit for the sample size will be
requested.

Selection

on | off

Default value

off

Hold at request

If this parameter is activated, then the run will be paused during the
request. If the parameter is deactivated, then the titration will be started in
the background.