5 control, Control – Metrohm 899 Coulometer User Manual

Page 53

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation

899 Coulometer

■■■■■■■■

43

5.5

Control

Menu

Control

In the dialog Control, the settings for the execution of a single determi-
nation or of one sample series are defined.

Sample table

If this parameter is activated, the sample data for a sample series can be
entered in a table (see Chapter 5.7, page 45).

Selection

on | off

Default value

off

Autostart

If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the number specified
has been reached (see Number of autostarts).

Selection

on | off

Default value

off

Note

If you are working with a sample changer with oven module (e.g. 885
Compact Oven SC
), then the Autostart parameter must be set to off,
because the sample changer starts the titration.

Number of autostarts

This parameter is visible only when Autostart = on.

Number of automatic starts.

Input range

1 ... 50

Selection

table

Default value

table

table
The number of automatic starts corresponds to the number of samples
in the sample table.