5 stop conditions, 6 calculation, 1 general – Metrohm 899 Coulometer User Manual

Page 91: Stop conditions, Calculation

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Parameters

899 Coulometer

■■■■■■■■

81

Electrode test

In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out as soon as the
determination is started.

Selection

on | off

Default value

off

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Input range

0.1 ... 999999.0 s

Default value

2.0 s

Temperature

Temperature entered manually.

Input range

–20.0 ... 150.0 °C

Default value

25.0 °C

7.1.5

Stop conditions

Menu

Parameters Stop conditions

The conditions for stopping a titration are defined under Stop condi-
tions
, if this does not occur automatically. This could be the case when
the endpoint set is not reached or if the stop criterion (see "Stop crite-
rion", page 78)
is not fulfilled.

Stop time

The titration is canceled when the specified time has elapsed since the
start of the titration.

Input range

1 ... 999999 s

Selection

off

Default value

off

7.1.6

Calculation

7.1.6.1

General

Menu

Parameters Calculation

A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations. Definition can be made for each calculation
as to whether the result is to be saved as a common variable.

Advertising