7.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

78

■■■■■■■■

899 Coulometer

Titration rate

slow

optimal

fast

Max. rate

1000.0 µg/min

maximum

maximum

Min. rate

0.3 µg/min

15.0 µg/min

30.0 µg/min

Stop criterion

The titration is canceled when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be canceled. The stop conditions (see Chapter 7.1.5,
page 81)
always lead to a stop, even if the stop criterion has not been
reached.

Selection

drift | time | rel. drift | off

Default value

rel. drift

drift
The titration is canceled when the stop drift has been reached.
time
The titration is canceled if the endpoint has been exceeded during a
certain time period (Delay time).
rel. drift
The titration is canceled when the sum of the drift at the start of the
titration and the relative stop drift has been reached.
off
The titration will not be canceled until the stop conditions have been
fulfilled.

Stop drift

This parameter can only be edited with Stop criterion = drift.

The titration is canceled when the endpoint and the stop drift have been
reached.

Input range

1 ... 999 µg/min

Default value

5 µg/min

Delay time

This parameter can only be edited with Stop criterion = time.

The titration will be canceled if the following criteria are both met simulta-
neously.

The voltage at the indicator electrode lies in the range "endpoint +5
mV and endpoint –2 mV".

The time defined here has expired.