4 methods, 1 method templates, 2 loading a method template – Metrohm 899 Coulometer User Manual

Page 50: Methods, Method templates, Loading a method template

Advertising
background image

5.4 Methods

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

899 Coulometer

Variable

Description

DDC

Time for drift correction

MIM

Initial measured value, i.e. measured value prior to the
processing of the start conditions

MIT

Initial temperature, i.e. temperature prior to the process-
ing of the start conditions

MCM

End measured value

MCT

End temperature

DD

Duration of the entire determination

5.4

Methods

5.4.1

Method templates

The 899 Coulometer contains method templates which are already config-
ured except for a few parameters.

The following method templates can be selected:

KFC

Coulometric Karl Fischer titration.

The blank value is not taken into account for the
calculation.

KFC-Blank

Coulometric Karl Fischer titration minus the
blank value.

Blank

Coulometric blank value determination.

The method templates differ only in their calculation formulas.

5.4.2

Loading a method template

Proceed as follows to load a method template:

1 Opening method templates

In the main dialog, select Method and press [OK].

The method table with the stored methods opens:

2 In the function bar, select New and press [OK].

The list with method templates opens: