8 reports, Reports – Metrohm 899 Coulometer User Manual

Page 96

Advertising
background image

7.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

86

■■■■■■■■

899 Coulometer

Statistics

If this function is activated, then statistics calculations will be carried out
for all of the defined results.

Selection

on | off

Default value

off

Number of determinations

The number of determinations that are carried out for the statistics calcu-
lations.

If an additional determination has to be added to the determination ser-
ies, because one determination has been incorrect, for example, then this
can be accomplished in the statistical overview (see Chapter 5.11, page
57)
.

Input range

2 ... 20

Default value

3

7.1.8

Reports

Menu

Parameters Reports

The reports that will be printed out automatically in connection with a
determination are defined under Reports.

Results

The result report contains the calculated results, endpoints, sample data,
etc.

Selection

on | off

Default value

off

Curve

Curve report. The width of the curve is defined in the system settings (see
"Graphics width", page 68)
.

Selection

on | off

Default value

off

Calculations/Statistics

Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes recalculation with an external pro-
gram possible (see Chapter 10.9, page 103). If Statistics has been activa-
ted, then the following data will be printed out as well:

Result and sample size of the individual determinations

Mean value as well as absolute and relative standard deviation

Advertising