Metrohm 899 Coulometer User Manual

Page 94

Advertising
background image

7.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

84

■■■■■■■■

899 Coulometer

Decimal places

Number of decimal places used to display the result.

Input range

0 ... 5

Default value

R1: 1 / R2-R5: 2

Result unit

The result unit is displayed and saved together with the result.

R1
Selection

% | mg | mg/piece | mg/g | mg/mL | ppm | µg |
User-defined

Default value

ppm

R2-R5
Selection

% | mg | mg/piece | mg/g | mg/mL | ppm | µg |
User-defined

Default value

%

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

Save as CV

The calculated result can be saved as a method-independent variable, a
so-called common variable. The result is then also available in other meth-
ods for calculations. If Statistics has been activated, then the current mean
value of the determination series will be saved.

Selection

on | off

Default value

off

Method "Blank"

Result name

The result name is the text which will be shown in the results display and
in the report.

Entry

12 characters

Default value

R1: Blank / R2-R5: empty

Calculation formula "R1"

EP1

R2…R5

Display of the calculation formula. A special editor is opened for the defi-
nition (see Chapter 5.3, page 39).

Advertising