7.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

899 Coulometer

off
No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual.

Drift for manual drift correction.

Input range

0.0 ... 99.9 µg/min

Default value

0.0 µg/min

Automatic start

If this parameter is activated, then the determination is automatically star-
ted when the measured voltage suddenly changes. This setting will be
ignored as long as the working medium has not been conditioned.

Selection

on | off

Default value

off

Note

If you use the automatic start function, then we recommend that the
start drift be reduced and that a stabilizing time be defined.

Recommendation: Start drift

≤ 10 µg/min / Stabilizing time ≥ 30 s

Threshold value

This parameter can only be edited with Automatic start = on.

If the change in voltage is higher than this value, then the determination
will automatically be started.

Input range

0 ... 999 mV

Default value

50 mV

Stabilizing time

Waiting time, during which the measured drift has to be smaller than the
start drift defined until Conditioning OK is displayed. The start drift is
defined under Conditioning

Start drift.

Input range

0 ... 999999 s

Default value

0 s

Cond. stop time

Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped as soon as the specified time has elapsed.

Input range

1 ... 999999 s