5 Operation

899 Coulometer

■■■■■■■■

55

5.9.3

Editing the live parameters

Certain method parameters can be edited while a determination is being
carried out. The only parameters that can be modified are those that can
be selected. Nevertheless, all of the parameters are visible. The modified
parameters are taken into account at once. If, for example, you increase
the parameter Extraction time while the extraction time is still running,
then the new value will be taken into account at once. If however you
change this parameter after the extraction time has already expired, then
the changed value will not be taken into account until the next determina-
tion.

Proceed as follows to edit the parameters:

1 Displaying the main dialog

Press [BACK].

The main dialog is displayed. The determination continues to run in
the background.

2 Opening the main menu

Select Menu and press [OK].

3 Editing the method parameters

Select the menu item Parameters and press [OK].

Change the desired parameters accordingly.

4 Displaying the live dialog

Select the menu item Live dialog in the main menu and press
[OK].

or

Press [BACK] in the main dialog.

The live dialog is displayed once again.