4 configuring external devices, Configuring external devices – Metrohm 899 Coulometer User Manual

Page 78

Advertising
background image

6.4 Configuring external devices

■■■■■■■■■■■■■■■■■■■■■■

68

■■■■■■■■

899 Coulometer

6.4

Configuring external devices

Menu

System External devices

PC/LIMS report

Specification of the storage location for the PC/LIMS report. The PC/LIMS
report is a machine-readable report with all of the important data for a
determination. It can be saved as follows:

as a TXT file on a USB flash drive.

to a LIMS via an RS-232 interface. The 6.2148.030 RS-232/USB Box is
required for this purpose.

Selection

COM2 | USB Stick

Default value

USB Stick

COM2
The report is sent via the serial COM2 interface. The interface parame-
ters set in the dialog COM2 settings are used (see "Editing the COM2
settings", page 70)
.
USB Stick
The report will be saved as a TXT file on the USB flash drive in the
folder pc_lims_report.

Printer

If a printer is connected, then the printer type needs to be defined here in
order for the reports to be printed out correctly.

The printers that have the designation ESC-POS are so-called POS printers
(point-of-sale printers), i.e. they print on continuous paper.

Selection

Citizen (ESC-POS) | Custom (ESC-POS) | Epson |
Epson (ESC-POS) | HP DeskJet | HP LaserJet |
Seiko (ESC-POS)

Default value

Custom (ESC-POS)

Graphics width

Adjust the width of the curve to be printed out to the paper width of the
printer to be used. The default value depends on the selected printer. The
height of the curve is 2/3 of the width.

Input range

100 ... 3000 Pixels

Keyboard layout

A commercially available USB keyboard can be connected to make it eas-
ier to enter text and numbers. Specify the country-specific keyboard lay-
out.