FUNCTIONAL DESCRIPTIONS AND SPECIFICATIONS

CHAPTER 2

Rear Panel (Input/Output)





Figure 2-5: SURETEST Rear Panel

A- SOLENOID CONTROL (Hirose RM15TRD-12S)

PIN FUNCTION
10

Loader Relay Common

11

Loader Relay CCW Limit

12

Loader Relay CW Limit


B

ANALOG OUTPUT (BNC)

Analog output provides a voltage level output that is directly proportional to transducer input,
from -1.8V (full scale counter clockwise) to +1.8V (full scale clockwise). Zero offset +/-100 mV.
Linearity +/-1% of reading. Maximum load = 5 mA.

When using the 4-in-1 transducer, zero output fluctuates until one of the transducers is
scan selected by applying 5% of full scale torque.

C-

Foot Pedal Switch Input


D-

Fuse Holder

Use AC fuse at specified rating only. ( 3.15 Amp).


E-

INPUT AC VOLTAGE

Automatic Selection of any AC voltage from 100VAC to 230VAC, 50-60 Hz.