CALIBRATION

CHAPTER 5

5-54

Use DOWN Cursor key to scroll down to Calibration Item. Press ‘ENTER’.

The display reads:

4.

Enter the following code into the display using the Cursor Keys:

5111

( Left / Right to select digit ; Up / Down = Increment / Decrement value ; ‘Enter’ to accept)



Calibrate tester

Using a laboratory millivolt meter, measure the voltage across pin 2 (ground) and 19 (Gage

Excitation) of the calibration fixture.


The

display

reads:


Example:

3.0211 volts


Program the measured voltage into the SURETEST Base Unit using the Cursor Keys. (Use Up,
Down, Left and Right Cursor Keys to match the Suretest display to the millivolt-meter).
Then Press ‘Enter’ to accept.

5.

Next, the display reads:

Short pins 5 and 6 together on the calibration fixture. Then hit Enter key.


6.

Next, the display reads:


Remove the short between pins 5 and 6 of the calibration fixture.