TORQUE CALIBRATION SYSTEM SETUP

CHAPTER 4

4-1

Clear Single Memory Location

Use the procedure in this section to clear the a single memory location.

1. Press (Data) to enter data list display. The display reads:


2.

Use the Up or Down Cursor to move to the current data. The units and present
memory location are given on the top line. The total number of memory locations
is to the right -- separated by a forward slash (/). The bar graph to the right of the
data list indicates how far in the list the current data display is positioned.

+04.456 in-lb, Time: 23:19 Date: Feb 10, 1999

3. Press

(DEL 1) to delete one data entry. The display automatically updates

(decrements by one) the total memory locations on the top line.



Clear All Memory Locations

Use the same above procedure for single memory location. Press (Clear) to

clear all

memory locations. The following warning appears:

Select “Yes” to clear all memory.

An audible alarm sounds momentarily. The SURETEST automatically returns to
measurements mode.


Setting Up Printer/PC Ports

The SURETEST is equipped with two RS-232 serial communications ports for use with
an external printer or personal computer. The transfer rate is programmable from 300 to
19.2K baud rate with 8 data bits, 1 stop bit and no parity. The recommended default is
19200 baud for Optional Port and 9600 for Serial Printer.

Communication setup as:

(300 – 19200); 8 Data bits; 1 Stop Bit; No Parity


Connect the Printer port to a computer or printer. All (serial) communications will take
place under User’s command(s) Edit Parameters set-up mode.