CALIBRATION

CHAPTER 5

5-51

Press F2 key and select the desired torque unit.

10.

Using the Up and Down Cursor keys, enter the Full Scale value for the 4-in-1 transducer to calibrate.
Possible values are:
1

=

50 in-lb

2

=

400 in-lb

3 = 1000

in-lb

4

=

250 ft-lb

( For 4 in 1 Transducers display shows : CHECKING TRANSDUCER.)
Apply load until “ Full Scale [units] shows on display.

11.

Press ENTER. To accept 50 in.lb Transducer.

Press any key to zerotare.


12.

Press any key. The LCD displays:


13.

Apply torque using the precision arm, hanger or certified weights listed as calibration point 1 in the

calibration check point table in this chapter. When the reading is stable, push and hold the Up or Down Cursor Keys
to get the correct torque value reading. Repeat the procedure for points 2,3,4, and 5 of the same table.

Apply 10% Torque = 5 in.lb