Printer serial output port, Serial communications, Analog output – CDI Torque SURETEST Torque Calibration System User Manual

Page 23: Loader control relays, Display capacity, Units of measurement, Accuracy, Range and resolution, Temperature

Advertising
background image

FUNCTIONAL DESCRIPTIONS AND SPECIFICATIONS

CHAPTER 2

Printer Serial Output Port

Default: 9600 Baud Rate.

Serial Communications

RS-232 (true)
300—19.2K Baud
8 data bits
1 stop bit
no parity


Computer Serial COM Port
(OPTIONAL)

Default: 19,200 Baud Rate.

RS-232 (True)

Analog Output

+(CW)/-(CCW) 1.8V at transducer full range linearity
±1% of reading

Loader Control Relays

Two, normally open, Form A, rated 12VDC @ 1/2A close contact at 110% CW or CCW of torque
transducer range. For additional information refer to Transducer and System specifications.

Display Capacity

(16 bit A/D), 5 digits, ±32,000 counts

Units of Measurement

Torque
ft-lb, in-lb, in-oz, Nm, dNm, cNm, mkg, cmkg


Torque Resolution

Refer to transducer range and display resolution tables in Functional Description of this chapter.

Accuracy

±0.2% of reading @ 25°C, within specified range, when used as prescribed with the 2000-600-02
Loader.

Range and Resolution

Refer to transducer range and display resolution tables in Functional Description of this chapter.

Temperature

Operating Temperature
10 to 32°C (50 to 90°F)
Storage
-20 to 50°C (-2 to 122°F)
Humidity
Up to 90% non-condensing
Temperature Drift
+0.02%/°C (+.011%/°F)