SETUP AND PROGRAMMING

CHAPTER 3

3-28

Programming Setup

Setting up Date and Time

1. Press (Setup).


2. Scroll down to select Clock Adjust. Press ‘Enter’.
The display reads:

3. Use Left or Right Cursor to select the date/time element to be changed.


4. Increment or decrement the date/time element by repeatedly pushing Up or Down

Cursor. Push and hold Up or Down Cursor to change the display quickly.


5. Press ENTER to update date and time and to return to measurements mode.


Setting Up High and Low Limits Alert

Use SET HIGH and SET LOW to protect the tool under test or serve as a fastener
installation torque preset alert. The SURETEST produces a constant audible tone when
the input torque/force exceeds the SET LOW limit. It then produces a pulsating tone
when the input torque/force exceeds the SET HIGH limit.

It is important to program these limits if statistical analysis is to draw a histogram which
includes reporting out-of-limit parameters.

Changing transducers does not automatically change SET HIGH and SET LOW limits.

Setting High or Low Torque Limits

1. Press (Setup).
2. Scroll down to select Edit limits.

The display reads: