Specifications, Suretest torque calibration system, System accuracy – CDI Torque SURETEST Torque Calibration System User Manual

Page 22: Temperature drift, Display accuracy, Temperature, Operating temperature, Storage, Dimensions, Weight

Advertising
background image

FUNCTIONAL DESCRIPTIONS AND SPECIFICATIONS

CHAPTER 2

Specifications

SURETEST Torque Calibration System

System Accuracy

±0.25% of reading @ 25°C
(Suretest and transducer calibrated together)

SURETEST Transducers used with, but not calibrated to, another SURETEST provide a system
accuracy of ±0.5% of reading @ 25°C.

Temperature Drift

+0.03%/°C (+0.017%/°F)


Display Accuracy

±0.05% of reading @ 25°C

Temperature

Operating Temperature

10 to 32°C (50 to 90°F)

Storage

-20 to 50°C (-2 to 122°F)
Humidity
Up to 90%, Non-condensing

Temperature Drift

+0.011%/°C, (+0.006%/°F)

Dimensions

Width 10.5“
Height 4.5"
Depth* 10.5"

*Includes handle, feet and connectors

Weight

8 .4 lbs


Fuse:

Amperage, 3.15 amps

Data Storage/Recall w/Date Time Stamp

3,000 measurements

Statistical Analysis

Max, Min, Range, Mean, Sigma N, Sigma, Cp, Cpk
%Error, -NoGo, +NoGo, Histogram.

Advertising