CALIBRATION

CHAPTER 5

5-55

Using a laboratory millivolt meter, measure the voltage across pins 5 and 6.


Example: 08.0622 mV

7.

Program the measured voltage into the SURETEST Indicator as demonstrated in the example,

“+08.0622” by pushing and holding the Up or Down Cursor keys.


8.

Press ENTER. The display momentarily reads:
SAVING...


9.

Turn the power to OFF. Remove the calibration fixture.