TO R Q U E C E L L C A L I B R A T I O N

Torque Cell Calibration

Version 2

WinPEP 7 User Guide

A-3

If you are only performing a Zero Calibration, click Finish.

Figure A-2: Calibration Mass Window

Enter the calibration number stamped near the bolt pattern at the end of the
calibration arm. If you do not have enough room to use the bolt pattern closest to
the end of the calibration arm, use the number stamped near the bolt pattern in
the center of the arm.
Note: Dynojet recommends you secure the calibration arm using the bolt pattern
closest to the end of the arm unless space constraints in your dyno room do not
allow you to.
Refer to one of the manuals from the list on page A-1 for more detailed
information on the calibration arm number.