TO R Q U E C E L L C A L I B R A T I O N

Torque Cell Calibration

Version 2

WinPEP 7 User Guide

A-5

While installing the calibration weights, you should notice the Torque Gauge on the
Torque Cell Calibration Window moving from 0 to about 500 foot-pounds.

Note: The Torque Gauge may or may not be in this range.
• If the torque cell has been previously calibrated incorrectly or has not been

calibrated for a while, the gauge may show values out of this range until

calibration is complete.

• If you use the bolt pattern in the center of the calibration arm, the gauge will

show values around 250-260 foot-pounds.

Note: Let the torque gauge needle stabilize before clicking Next.

9

From the Span Calibration window (Figure A-3), click Next to continue.
At this point, the value on the gauge should match the value on the calibration
arm.

10 Click Previous to calibrate a different torque cell. Repeat steps 7 through 10.
11 Once you have completed calibrating the torque cell(s), remove the calibration

arm and weights and click Finish.

Figure A-4: Calibration Is Complete Window