PARTS COUNTING-CHECK
This feature permits establishing a set quantity of pieces as a
criteria for similar items that can be quickly checked against the
sample. In the Parts Counting Check mode, the displayed data
includes, present count (Pcs), APW, Sample Size, Tare, Weight
and a bar graph indicating UNDER, ACCEPT (blinking) and
OVER.

Enter the PARTS COUNTING SETUP menu and change Mode to
Check.

The balance is now in the PC CHECK mode.

PARTS COUNTING

01/01/03 03:05:45 PM

SAMPLE SIZE

SETUP

NEW COUNT

Add 10 Pieces, Press Enter

0.00 g

APW: 0.00g
Sample Size: 10

Under: 5

Tare: 0.00g

Over: 10

The balance indicates to add 10 pieces and has an Under value
of 5 and an Over value of 10 set up as default values and an
average piece weight has not been set. Place 10 pieces on the
pan and press the Enter button.

An average piece weight has been set by placing the pieces on
the pan. To use the PC Check, the Over and Under values must
be set as well as other display setup parameters. Refer to
adjustments and set the balance parameters to fit specific
requirements.

ADJUSTMENTS

The PC Check display contains three soft keys at the top of the
screen SAMPLE SIZE, SETUP and NEW COUNT.

NEW COUNT
When NEW COUNT soft key is selected, follow the screen
instructions “Add 10 Pieces, Press Enter “. After pressing the
Enter button, the balance calculates the APW and the display
indicates the number of pieces.

SETUP
Using the left arrow key, select SETUP soft key and press the
Enter button. The PARTS COUNTING SETUP is shown.

Two additional entries have been added to the PARTS COUNTING
SETUP, they are Over and Under. Refer to the Adjustments section
of parts counting for a description of all other settings.

Over 10: 0-9999 (default is 10)
Defines the over limit in pieces.

Under 5: 0-9999 (default is 5)
Defines the under limit in pieces.

Display Settings: Custom, Default (default is Default)
Selecting Display Settings to Custom will bring up the DISPLAY
SETTINGS screen that will allow individual display items to be
turned on or off. Sample Size, Tare, Weight are described under
parts counting Display Settings.

PARTS COUNTING SETUP

Library

:

PARTS COUNTING

Mode

:

Check

Unit

: g

Sample Size

: 10

APW

: Off

Tare

: 0.00

Auto Optimize

: Off

Over

: 10

Under

: 5

Display Settings

: Custom

Size/APW Key

: Unlocked

New Count Key

: Unlocked

PARTS COUNTING 01/01/03 03:05:45 PM

SAMPLE SIZE

SETUP

NEW COUNT

20.00 g

APW: 2.00g
Sample Size: 10

Under: 5

Tare: 0.00g

Over: 10

10