Check weighing, Under, 6 filling – Rice Lake Ohaus Explorer Pro High Capacity Instuction Manual User Manual

Page 24

Advertising
background image

EN-20

Explorer

Pro

CHECK WEIGHING 01/01/03 03:05:45 PM

UNITS

SETUP

CALIBRATE

Over: 15.00 g
Target: 10.00 g

Diff: -10.00 g

Under: 5.00 g

Diff: -100.00 %

UNDER

0.00g

*

Exit:
When selected, returns to Check Weighing mode.

UNITS
When the UNITS soft key is highlighted, each press of the Enter
button cycles through the measuring units that are On.

Library Name:
Allows storage and recovery of parameters and results. When
a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 18 characters.

Unit: g, kg, etc...(default is g)
Allows the selection of all activated units.

Over: 0-99999999 (default is 15)
Defines over limit value.

Target: 0-99999999 (default is 10).
Defines target value.

Under: 0-99999999 (default is 5)
Defines under limit value.

Display Settings: Custom, Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Check Weighing mode can be individually turned on or off. If
Display Setting is set to default, all the information is turned
on.

Reading allows a choice of either weight readings to be
displayed in large numeral or message that displays OVER,
ACCEPT and UNDER in large letters. Display below indicates
message format.

3.4.6 Filling

Filling permits entering a target reference weight. Material can
be placed on the pan and by monitoring the Difference weight
and the bar display, an accurate fill can be achieved. A split
bar display indicates 0% to 90% on the first bar and 90% to
110% on the second bar.

FILLING 01/01/03 03:05:45 PM

TARGET

SETUP

NEW TARGET

0.00 g

Target: 50.00 g
Diff: -50.00 g
Pct: 0%

0.00

g

*

0%

50%

90%

110%

100%

USE

FILLING
A sample (target weight) is placed on the pan and the Enter
button is pressed. The balance stores this weight as shown on
the display as Target. The sample is removed from the pan and
the material is added to the pan. The display provides Target
Weight, Difference weight, Percentage and the split bar at the
bottom of the screen displays the percentage of the sample
compared against the target weight. When a container is used,
zero the balance before placing the actual sample on the pan.
See Adjustments to set other balance parameters.

ADJUSTMENTS

The FILLING display contains three soft keys at the top of the
screen TARGET, SETUP and NEW TARGET.

NEW TARGET
With the NEW TARGET soft key highlighted, press the Enter button.
Follow display instructions to “Add Target Weight, Press Enter”.
This establishes a new target weight and is displayed on the
screen.

SETUP
When SETUP soft Key is highlighted, press the Enter button. The
FILLING SETUP display is shown.