3.5 Balance Settings

The Explorer Pro balance contains eleven submenus that
are accessible from the Main Menu. The submenus are
Calibration, Balance Options, Readout, Application Modes,
Units, Interface, Print Options, GLP Print Options, Library,
Lockout and Factory Reset.

Each of the submenus contain settings that will affect the operation
of the balance. Please review all submenus settings to obtain
the best performance from the balance. Make the necessary
settings to suit specific needs.

3.5.1 Calibration

Press the Menu button and select CALIBRATION. Press the Enter

Enter

Enter

Enter

Enter

button, the CALIBRATION screen is displayed. Explorer Pro
balances offers a choice of five calibration methods: Internal
Calibration, Autocal Calibration, Span Calibration, Linearity
Calibration and User Calibration,

Linearity, Span and User calibration are disabled for Type
Approved/LFT balances Class II and Class III.

AutoCal: On, Off (default is On)
When AutoCal is set ON, the balance performs a self calibration
when a measured predefined delta temperature change occurs.
After calibration, the display returns to the last application mode.
Environmental conditions must be met for the AutoCal to be
successfully completed.

AutoCal Delta: -100 - +100 (default is 0)
AutoCal Delta allows the internal calibration mass value to be

adjusted. This permits calibrating the balance using an external

mass which is traceable to a certified standard.

Perform the internal calibration procedure. After the calibration is

completed, zero the balance.

Place a certified mass equal to the

span calibration value

span calibration value

span calibration value

span calibration value

span calibration value

of the balance.

Compare the reading on the balance to the expected weight
being used. If the reading is above the expected weight, the
delta entered in digits is negative. If the reading is below the
expected weight, the delta entered in digits is positive.
See following example:

Actual Weight Reading:

200.0014

Expected Weight Reading: 200.0000
Delta Weight (d):

0.0014

Delta weight in Digits:

-14

Recalibrate using internal calibration. After calibration, place

the certified mass on the pan and see if the mass agrees with

the displayed value. If not, repeat procedure until internal

calibration reading agrees with the certified mass.

Span Calibration:
Span calibration utilizes two calibration points, one at zero
load and the other at specified full span. See the following
table.

IMPORTANT !

DO NOT DISTURB THE BALANCE

DURING CALIBRATION.

CALIBRATION

Internal

AutoCal

: Off

AutoCal Delta (d): 0
Span
Linearity
User
User Cal Weight (g): 500.00
Cal Test
Exit

Internal Calibration
Calibration is accomplished using the internal calibration mass.
Internal calibration can be performed at any time providing the
balance has warmed up to operating temperature and is level.