Custom unit (cst), Hardware lockswitch, Lft default table – Rice Lake Ohaus Explorer Pro High Capacity Instuction Manual User Manual

Page 47: Printing data, Switch plate, 10 lockout, 11 factory reset

Advertising
background image

Explorer

Pro

EN-43

3.5.10 Lockout

Provides the capability to lock the Menu settings individually to
protect selected parameters against modifications. Locked means
the items can be viewed but not changed.

Press the Menu button, and select Lockout.

Press the Enter button LOCKOUT screen is displayed.

LOCKOUT

Calibration

:

Unlocked

Balance Options

: Unlocked

Readout

: Unlocked

Application Setup

: Unlocked

Application Modes

: Unlocked

Units

: Unlocked

Interface

: Unlocked

Print Options

: Unlocked

GLP Print Options

: Unlocked

Factory Reset

: Unlocked

Exit

FACTORY RESET

Calibration

:

No

Balance Options

: No

Readout

: No

Application Modes

: No

Units

: No

Interface

: No

Print Options

: No

GLP Print Options

: No

Lockout

: No

Global Reset

: No

Accept Changes

: No

Balance Options:
Language: English
Display Date/Time: Date/Time
Date Format: MM/DD/YY
Time Format: 12 Hour

Readout:
Filter Level: Medium
Stability Level (d): 1.0
Auto Zero (d): 0.5
Auto Tare: Off
Legal for Trade: Off

Application Modes:
Weighing: On
All others Off

Units:
Grams On, all others Off.

Interface:
Baud Rate: 2400
Parity: None
Data Bits: 7
Stop Bits: 2

Print Options:
Auto Print: Off
Print Interval: Off
Stable Data: Off
Numeric Data: Off

GLP Print Options:
Balance ID=OHAUS
User Name = OHAUS
All others Off

Lockout:
All Unlocked

The default setting for all items on the LOCKOUT screen are
Unlocked.

3.5.11 Factory Reset

The FACTORY RESET allows parameters to be set to the default
values. Each menu item can be reset by setting to Yes. After
accepting the changes, the balance will reset the selected menus
items. Global Reset will change all menus to a factory setting in
one step.

Press the Menu button, and select Factory Reset.

Press the Enter button FACTORY RESET screen is displayed.

The balance default parameters are listed as follows:
Calibration:
Auto Cal: On
Autocal delta: 0
User Cal weight = Span weight

Advertising