3 readout, 4 application modes, 5 units – Rice Lake Ohaus Explorer Pro High Capacity Instuction Manual User Manual

Page 43

Advertising
background image

Explorer

Pro

EN-39

Brightness:
Allows adjusting the brightness level of the display.

Audible Signal: On, Off (default is Off)
With audible signal set On, each button press will generate a
tone.

Exit:
When selected, returns to previous screen.

3.5.3 Readout

Press the Menu button, and select Readout.

Press the Enter button. The READOUT screen is displayed.

READOUT

Filter Level

:

Medium

Stability Level (d)

: 1.0

Auto Zero (d)

: 0.5

Auto Tare

: Off

Legal for Trade

: Off

Exit

Filter Level: Low, Medium, High (default is Medium)
Sets the balance reading averaging level to a value of Low,
Medium or High. A higher filter level will provide a more repeatable
measurement.

Stability Level: 0.5, 1.0, etc... (default is 1.0)
Sets the balance stability level for the stability indicator to either
0.5, 1.0, 2.0 or 5.0. A 0.5 setting is equivalent to end result of
.5 display digit. A 5.0 setting is equivalent to end result of 5
display digits.

Auto Zero: Off, 0.5, etc... (default is 0.5)
Sets the balance auto zero level to either Off, 0.5, 1.0, 2.0 or
5.0. Auto Zero tracking compensates for the drift at zero load.
0.5d means 0.5 display digit drift compensation per second.
5.0d means 5 display digit drift compensation per second.

Auto Tare: On, Off (default is Off)
Sets the Auto Tare function to On or Off. This feature provides
auto tare in all application modes. When Auto Tare is On, the
balance waits for the container/load to be placed on the pan
and automatically tares it. This function is repeated for each
new container/load.

Legal for Trade: On, Off (default is Off)
Sets the Legal for Trade (LFT) function either On or Off. Refer to
paragraph paragraph 3.5.12 for additional information.

3.5.4 Application Modes

Press the Menu button, and select Application Modes.

Press the Enter button APPLICATION MODES screen is displayed.

All application modes are displayed on this screen. Each mode
may be set on or off. These modes are cycled through when the
Mode button is pressed.

3.5.5 Units

Press the Menu button, and select Units. The UNITS screen
displays a list of the available weighing units. Depending on
the model, some units are not available.

APPLICATION MODES

Weighing

:

On

Parts Counting

: Off

Animal Weighing

: Off

Percent Weighing

: Off

Check Weighing

: Off

G/N/T

: Off

Filling

: Off

Differential

: Off

Formulation

: Off

Statistics

: Off

SQC

: Off

Density

: Off

UNITS

Milligram (mg)

:

Off

Gram

: On

Kilogram

: Off

Oz avdp (oz)

: Off

Pound avdp (lb)

: Off

Carat (ct)

: Off

Penny weight (dwt)

: Off

Troy OZ (ozt)

: Off

Grain (GN)

: Off

HK Tael (Hkt)

: Off

SG Tael (SGt)

: Off

ROC Tael (ROt)

: Off

Scroll down using arrow button to view remaining units.

Advertising