Instruction Manual - TORBAL AD Series

7.1.1.1. Setup

To change the calibration mass setting, follow the steps below;

1. Press the “M” key to enter the Main Menu

2. When command CALIB is displayed press the “T” (YES) key,

3. The scale will display commands Start and Setup sequentially. Press the “T” key when Setup is

displayed.

4. The current calibration mass setting will be displayed. To change the calibration mass, press the 0

(NO) key. After making the selection press the “T” (Yes) key to accept and confirm the setting. The

command DONE will be displayed, indicating that the setting has been saved.