Torbal AD50 Series Manual User Manual

Page 16

Advertising
background image

Instruction Manual - TORBAL AD Series

6. Once the scale tares, LOAD will be displayed. At any time during the LOAD command, place a

single calibration weight in the middle of the pan.

7. When the calibration weight is placed on the pan, the scale will automatically calibrate. The display

will read “C” while calibrating. Because calibration is a very sensitive process, make sure the scale

is not exposed to any air movements or vibrations while it is calibrating.

8. Calibration can last from 15 to 30 seconds. When it is finished the display will read CAL Fin and the

scale will automatically return to the weighing mode. Once the CAL Fin command disappears and

the scale returns to weighing, remove the weight from the pan.

16

Advertising