Instruction Manual - TORBAL AD Series

7.1.1.2. Start

To calibrate the scale, follow the steps below:

1. Press “M” to enter the Main Menu

2. When CALIB is displayed, press “T” (YES).

3. The scale will display commands Start and Setup sequentially. Press , “T” , when Start is displayed

.

4. The scale will display CAL, indicating that the calibration mode has been entered.

5. Before calibration begins, the scale will perform automatic tare. Taring will be displayed during auto

tare.