Instruction Manual - TORBAL AD Series

4. After enabling the totalizing function, DONE will be displayed, indicating that the function has

been configured and it is ready to use. To use the function follow the directions in Chapter 8.

To exit the mode menu, use the “C” key.

7.1.4.Automatic Zero Setting Mechanism (AZSM)

All AD scales are equipped with AZSM, the “Auto Zero Setting Mechanism.” AZSM

automatically maintains a “center of zero” condition within +/- 6d (i.e. 6 mg on scales with 0.001

readability). To enable or disable AZSM follow the steps below.

1. Enter the Main Menu by pressing the “M” key.

29

Table 7.1.3.F: Totalizing Operating Modes

Automatic

Totalizing and Adding: Automatic

When set to Automatic, the scale will automatically

add the first stable weight reading to the total.

Manual

Totalizing and Adding: Manual

When set to Manual each weigh has to be manually

added to the total by pressing a designated key.