Instruction Manual - TORBAL AD Series

5. Place the sample on the pan. The weight of the sample will be displayed.

Note: The reference sample must be equal to or greater than 100d. Attempting to set a sample lower

than 100d will result in a SAPL LO error.

6. Once the weight of the sample has stabilized and the stability indicator has been displayed, press

“T” to accept the sample. Once a new sample weight is set it will be saved until changed or

cleared through the master RESET.

7. After setting the sample weight, the command READY will be displayed. Remove the sample and

place the object on the pan. The scale will express the weight of the object as a percentage of the

stored sample.