Torbal AD50 Series Manual User Manual

Page 26

Advertising
background image

Instruction Manual - TORBAL AD Series

4. After enabling the display hold function, DONE will be displayed, indicating that the function has

been configured and it is ready to use. To use the function follow the directions in Chapter 8. To exit

the mode menu,, use the “C” key

26

Table 7.1.3.D: Display Hold Operating Modes

Automatic

Display Hold initialization: Automatic

Result Clearing: Automatic

Tare: Automatic

When set to Automatic, the scale will automatically

lock and hold the first stable weight reading on the

display. When the weight is removed the scale will

automatically clear the result. Once the result is

cleared the scale will automatically tare and prepare

for the next weighing.

Semi-Automatic

Display Hold initialization: Automatic

Result Clearing: Manual

Tare: Manual

When Set to Semi-Automatic, the scale will

automatically lock and hold the first stable weight

reading on the display. When the weight is removed

it must be manually cleared. Before starting the next

weighing the scale must be manually tared.

Manual

Display Hold initialization: Manual

Result Clearing: Manual

Tare: Manual

When set to manual, the first stable weight reading

on the display must be locked manually. When the

weight is removed it must be manually cleared.

Before starting the next weighing the scale must be

manually tared.