Chapter 5 - setup – Hardy HI 200DNWM DeviceNet Weigh Module User Manual

Page 35

Advertising
background image

CHAPTER 5 - SETUP

5-1

CHAPTER 5 - SETUP

SCOPE

All information contained in Chapter 5 per-
tains to software or firmware settings or pro-
cedures to prepare the HI 200DNWM weigh
module for calibration and operation. Alter-
natives to these procedures either explicit or
implied, contained in this section are not rec-
ommended. It is very important that the user
and service personnel be familiar with the
procedures contained in this chapter, before
going through the setup procedures.

Saving to Non-Vol-
atile Ram

NOTE:

If you have an instrument with version 1.5, please
refer to Devicenet Manual, REV B-6 or earlier for
instructions about Saving to Non-Volatile Ram or
contact Hardy Instruments Customer Service Depart-
ment.

In version 2.1, the HI 200DNWM does not
automatically save parameters to non-vola-
tile memory. To save parameters to non-vol-
atile RAM, set parameter 38 (Save non-
volatile command) to 1.

This request should be sent after changing
any configuration parameter. It does NOT
need to be sent after altering a command
parameters like TARE or ZERO. It does not
need to be sent after doing a calibration. The
necessary calibration data values are saved
automatically. However, if you change
parameters used during calibration, like
“number of averages”, or “WAVER-

Advertising