HI 200DNWM MANUAL

6-4

Step 6. Enter the Span Weight Value.
Step 7. Click on Cal LO.
Step 8. Enter the Cal LO weight.
Step 9. Click on Cal LO command.
Step 10. Set the Cal LO to “1”.
Step 11. Click on the Apply button.
Step 12. Click on Yes.
Step 13. Place the test weight on the

scale.

Step 14. Click on Cal HI command.
Step 15. Set the Cal HI command to “1”
Step 16. Click on the Apply button.
Step 17. Click on Yes.
Step 18. Click on Cmd Save non-vol.
Step 19. Set the Cmd Save non-vol to

“1”.

Step 20. Click on the Apply button.
Step 21. Click on Yes.
Step 22. The calibration is complete.