3.3

Power-Up

Before switching the device on,
remove the protective cover from
the haze ports. To switch on the
equipment, press the power switch
on the left side of the device.

The device will run a self-diagnostic
test after it is switched on, the
progress of which is indicated on
the display.

The first thing that appears on the
screen is the company logo along
with the software version number
and the copyright note. The version
number and year of copyright are
important for Customer Service if
you ever need to report a problem
with the device.

The device then carries out a self-
diagnostic test. The lamp, chopper,
and shutter are checked. If the
values for any of these components
are outside the tolerance range, an
error message is displayed.

If the lamp is faulty, the self-
diagnostic test is stopped. If this
happens, replace the lamp (see
Chapter 12). If an error is reported
for other components, please
contact our Customer Service.