Change Cal. Values
Input Trans Std.
Save Std Values
Escape:
SET UP
Initialize Instr.
Old New
100 90
Save Data: OPERATE
Escape: SET UP
Continuous Storage
To permanently store the altered
value of a calibration standard in the
device, go to the Change Cal.
Values. menu and select the
option “Save Std Values”
using the cursor buttons and
then press the “operate”
button.
The “Save Std Values” display
appears.
The old and new value are
presented here.
To confirm the new value, press the
“operate” button.
The device starts calibrating. After
the dark and light measurement,
you will be requested to place the
standard with the value just entered
at the haze port. Insert the standard
and press the “operate” button. The
calibration is finished and the new
transmission value is now
permanently stored.
Standards must be clean and
undamaged!