51
English
No statistic
values stored
SET UP
Printer
not ready
Escape:
Invalid date
Cause
The current statistics memory in the
measurement device does not
contain any data.
Cause
The connected printer is not re-
sponding. This error only occurs
when statistics data is being prin-
ted. The printer is not polled
during online mode.
Remedy
• Check that the printer cable is
correctly connected!
• Is the printer switched on?
• Is there paper in the printer?
• Is the printer expecting a
postscript file?
Cause
A date earlier than 1980 was input
during SET-UP.
Remedy
The date input was ignored and the
old date retained. Enter the correct
date.
16 Error Messages