Zero Calibration

Please cover

haze-port

Ready:

OPERATE

Escape:

SET UP

CALIBRATION

ASTM

6.1

Calibration

Thanks to the use of state-of-the-art
technology and the reference beam
principle, calibration does not have
to be carried out very often. We
recommend a calibration interval of
two months. Calibration is also
necessary:

• at the initial startup,

• in the event of a drastic change in

ambient temperature (for
example, if the device has been
moved to a new location),

• if a message appears on the

display requesting calibration (e.g.
following the modification of
stored calibration values),

• if the measurement method has

been changed.

6.2

How to Calibrate the Device

Calibration is carried out in the SET-
UP Menu. This menu is opened by
pressing the “set up” button. Using
the cursor buttons, select the menu
item “Calibration” and then press
the “operate” button. The dark
measurement is carried out. Cover
the “haze-port” with the dark
standard and press the “operate”
button again.

To interrupt the calibration procedu-
re at any time, press the “set up”
button.

6 Calibration