ALTANA Haze-Gard Dual User Manual

Page 53

Advertising
background image

52

English

Ref. Measurement

only in

Continuous-Mode

Memory

full

Adjusting

Calibration Values

Continue:

OPERATE

Online Mode

Keyboard inactive

Cause

The “reference” button was pressed
although the device is not in
Continuous Mode.

Remedy

None, this is merely provided for
information purposes.

Cause

The current statistics memory is full.

Remedy

Delete the last value or entire
memory by pressing the “delete
last” or “delete all” button.

Cause

The RAM and EEPROM contain
different calibration data as these
were not saved following the chan-
ging of calibration values.

Remedy

The calibration data are auto-
matically transmitted from the
EEPROM to the RAM and thus the
calibration values updated again.

Cause

The measurement device is in
online mode and is hence controlled
from the PC- Only the “operate” and
referencee buttons and the “foot
switch” are released.

16 Error Messages