Specimen

Measurement

Calibration

No compensation: different sphere efficiency

Specimen

Measurement

Calibration

With compensation: same sphere efficiency

ASTM D 1003

Measurement conditions during calibration and actual measurement are
different. While part of the light that enters through the outlet escapes
calibration, during measurement the outlet with the specimen is closed.
Thus, the light in the sphere is increased by the proportion which is reflected
on the specimen.

ISO 13468

Measurement conditions during calibration and actual measurement are the
same. The specimen is placed at the compensation outlet during calibration
and during measurement it is placed at the measurement outlet. This means
that the sphere efficacy is completely dependent on the reflectance
properties of the specimen. The difference between the two measurement
methods can be up to 2% for clear glossy specimens.