ALTANA Haze-Gard Dual User Manual

Page 40

Advertising
background image

39

English

T.Std. = 100

Confirm: OPERATE

Escape: SET UP

1000

Change Cal. Values
Input Trans Std.

Save Std Values

Escape:

SET UP

Adjusting

Calibration Values

Continue:

OPERATE

The value to be changed is shown
as a four-digit figure on the display.
The last digit here is the digit after
the decimal point.

Select the decimal
point to be changed
using the buttons:

and change the value using
the “up/down” cursor buttons.

Confirm the change by pres-
sing the “operate” button. The
device will return to the Change Cal
Values menu.

Temporary Storage

If you leave the Change Cal Values
by pressing the “set-up” button (i.e.
cancel), the altered value is adop-
ted temporarily. The device starts
calibrating. After the dark and light
measurement, you will be requested
to place the standard with the value
just entered at the haze port. Insert
the standard and press the
“operate” button. The calibration is
finished and the new transmission
value is now saved until the device
is switched off.

When the device is switched on
again, the following message
appears beside the company logo.

Calibration data is being updated.

Standards must be clean and
undamaged!

10 SET-UP