ALTANA Haze-Gard Dual User Manual

Page 30

Advertising
background image

29

English

reference

OPERATE

Transmittance

Measurement

Please place

Sample at

haze-port

Ready:

ISO

H n= 11

m= 14.6

s= 1.16

15.8

CONT. MODE

To add a value to the statistics,
press the “operate” button. The
following message appears on the
display:

Measurement active.

The device will return to continuous
mode. If the statistic function is acti-
vated (Chapter 8.3), the counter will
increase by one after measurement.

If you change the material to be
measured, a new compensation
measurement is required. Indicate
the change by pressing the “refer-
ence” button.

You will be asked to place the spe-

pensation measurement is triggered
by pressing the “operate” button.
The device will then return to conti-
nuous mode.

Haze

The haze measurement requires the
value of the total transmittance as a
reference for the measured speci-
men. This is indicated in a message
on the display. Place the specimen
whose total transmittance is to be
defined at the haze-port.

The device will return to Continuous
Mode on completion.

OPERATE

Measurement

Place

Sample at

comp-port

Ready:

Compensation

cimen in the comp-port. The com-

9 Continuous Mode