SSi Manual #4563 Rev. D Page 25 Multi-Gas Model 6000

flow until the values stabilize, which should be within one to two minutes. Enter the exact quantity of
Hydrogen in the calibration gas in the Hydrogen Span Value area, press the down arrow key once to
highlight the Start display, and press Enter. At the completion of the span calibration, the Hydrogen
Reading should be equal to the Span Value.

Example – The following will show how to calibrate the zero value with a target value of 0, and to calibrate

the span value with a target value of 39.99. When the screen first displays, the hydrogen zero value field is

already displayed. Enter a 0 and press Enter (Note – the screen will display 0.00). After the zero gas has

come to equilibrium, press the down arrow key once to highlight the Start display and press Enter. The

MGA will now zero calibrate the H2 cell. When the calibration begins, a message will be displayed below the

“Hydrogen Span Value” line that says “Processing”. If the zero calibration was successful, this message will

change to “Z1 Pass”. If the zero calibration was not successful, the message will change to “Z1 Fail”. Next,

switch the flow of gas from zero to span and allow the values to come to equilibrium. Press the down arrow

key once to highlight the hydrogen span value field. Enter a 39.99 and press Enter. Press the down arrow

key once to highlight the Start display and press Enter. The MGA will now span calibrate the H2 cell.

When the calibration begins, a message will be displayed below the “Hydrogen Span Value” line that says

“Processing”. If the span calibration was successful, this message will change to “S1 Pass”. If the span

calibration was not successful, the message will change to “S1 Fail”.

Auto Sequence Setup – Menu Page 28

This option is not yet implemented.

Gas or CV Configuration – Menu Page 29

This menu option requires the Super Systems Inc special passcode to view. The menu option is used in
setting up the individual gasses for the MGA and it should not be modified without contacting Super Systems
Inc at 800-666-4330. These gasses are preset at the factory and should not require any modifications.