Super Systems MGA6000 User Manual

Page 36

Advertising
background image

SSi Manual #4563 Rev. D Page 36 Multi-Gas Model 6000

1819

Command to send to cell

1820

sensor start up timer (500ms increments)

1821

read SPI

1822

SPI data counter, high byte

1823

SPI store state 0 = normal operation (store)

1 = receiving data from boss, 2 = OK to send to Cell, 4 = sending to cell

1824

timer on SPI response, 10ms inc

1825

indicates

SPI

responded

1826

Last

command

used

1827

Result of the last command

* Registers 1828 to 1829 not used

* Data sent by IR cell in master mode

1830

decimal place //debug

1831

gas reading

* calc structure
1832

deg C (1832 - 1833)

1834

volts (1834 - 1835)

1836

atten (1836 - 1837)

1838

zero_volts_tc (1838 - 1839)

1840

span_volts_tc (1840 - 1841)

1842

atten_tc (1842 - 1843)

1844

curve_in (1844 - 1845)

* cal structure

1846

degCZ (1846 - 1847)

1848

degCS (1848 - 1849)

1850

Zero_volts (1850 - 1851)

1852

Span_Volts (1852 - 1853)

1854

ztc (1854 - 1855)

1856

stc (1856 - 1857)

1858

zero_offset (1858 - 1859)

1860

span_target (1860 - 1861)

1862

Command

1863

Command

result

* curve structure

1864

c_dummy (1864 - 1865)

1866

coeff_0 (1866 - 1867)

1868

coeff_1 (1868 - 1869)

1870

coeff_2 (1870 - 1871)

1872

coeff_3 (1872 - 1873)

1874

coeff_4 (1874 - 1875)

1876

coeff_5 (1876 - 1877)

1878

coeff_6 (1878 - 1879)

1880

coeff_7 (1880 - 1881)

* EEROM structure

1882

software revision (1882 - 1890)

* Define date_of_cal

1891

Filter

1892

span_offset (1892 - 1893)

* Registers 1894 to 1898 not used

* calculated values

1899

gas_value_adj