SSi Manual #4563 Rev. D Page 33 Multi-Gas Model 6000

722

PIC 3 IR cell link to gas n

723

PIC 4 IR cell link to gas n

* Register 724 not used
725

PIC 1 span gas value

726

PIC 2 span gas value

727

PIC 3 span gas value

728

PIC 4 span gas value

* Registers 729 to 730 not used

731

IR shim factor

732

Pressure compensation factor; 0 = off, 10 = 1.0, max 2.0

733

CH4 factor

734

IR CO compensation factor

735

Control output assignment 1

736

Control output assignment 1

737

Control output assignment 1

738

Control output assignment 1

739

Calculation flag 0 = COF, 1 = MMI PF

740

pump ON delay timer

741

sample OFF delay timer

742

sample ON delay timer

743

Interval timer for COF/PF adjust

744

pump mode; 0 = off, 1 = on, 2 = auto

* Registers 745 to 749 not used

750

Dac 1 assignment

751

Dac 2 assignment

752

Dac 3 assignment

753

Dac 4 assignment

754

Dac 1 zero

755

Dac 1 span

756

Dac 2 zero

757

Dac 2 span

758

Dac 3 zero

759

Dac 3 span

760

Dac 4 zero

761

Dac 4 span

762

selected sequence storage (762 - 777)

4

sequence

numbers

per

integer

* Registers 778 to 799 not used
800

Gas1/Gas2 ratio

* Registers 801 to 819 not used

820

HiTech CO compensation ratio * 1000

821

HiTech CO2 compensation ratio * 1000

822

HiTech CH4 compensation ratio * 1000

* Registers 823 to 824 not used
825

Calculated new factor for COF/PF adjust

826

pump control; 0 = off, 1 = on

* Registers 827 to 837 not used

838

calculate p trim value (need 92)

839

calibration timer

840

calibration inhibit (IN/OUT of range)

bits 0,8 = PIC 1 gas, bits 1,9 = PIC 2 gas, bits 4,12 = PIC 3 gas

bits

5,13

=

PIC

4

gas

841

calibration function

* Registers 842 to 878 not used